Antistasis Retrograde Flow Vascular Catheter: A Novel Solution to Thrombogenicity: A Computational Fluid Dynamics Study.

2020 
BACKGROUND: Catheter-related thrombosis (CRT) is a serious complication of vascular catheters. Retrograde catheter insertion has been shown to decrease pericatheter hemostasis and thrombosis, but it is technically challenging. The current in silico trial is an analytical approach to evaluating different approaches to designing retrograde flow into a vascular catheter. METHODS: The novel catheter design aims to provide antistasis retrograde flow (ASRF) of fluid through multiple backward-directed side openings, with a self-closing terminal opening to facilitate standard insertion. Four different models of the catheter were evaluated by computational fluid dynamic studies, with retrograde-angled openings of 15 degrees , 30 degrees , 45 degrees , and 60 degrees to the long axis of the catheter. RESULTS: ASRF successfully reduced the areas of fluid stagnation in models with 15 degrees and 30 degrees openings. Models with 45 degrees and 60 degrees did not significantly reduce stagnation. ASRF is reversed by the main bloodstream after a few millimeters. The novel catheter design achieved a slightly higher saline flow rate compared with the standard catheter (89.75, 91.72, 94.13, and 94.26 mL/min for 15 degrees , 30 degrees , 45 degrees , and 60 degrees designs, respectively, versus 86.93 mL/min for the standard catheter). CONCLUSIONS: The novel ASRF vascular catheter reduces pericatheter fluid stasis and has the potential to reduce CRT. Further in vitro and in vivo trials are warranted to validate these findings and evaluate clinical efficacy.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    26
    References
    0
    Citations
    NaN
    KQI
    []