Digital Breast Tomosynthesis: Observer Performance Study
2009
OBJECTIVE. The purpose of this study was to compare in a retrospective observer study the diagnostic performance of full-field digital mammography (FFDM) with that of digital breast tomosynthesis.MATERIALS AND METHODS. Eight experienced radiologists interpreted images from 125 selected examinations, 35 with verified findings of cancer and 90 with no finding of cancer. The four display conditions included FFDM alone, 11 low-dose projections, reconstructed digital breast tomosynthesis images, and a combined display mode of FFDM and digital breast tomosynthesis images. Observers rated examinations using the screening BI-RADS rating scale and the free-response receiver operating characteristic paradigm. Observer performance levels were measured as the proportion of examinations prompting recall of patients for further diagnostic evaluation. The results were presented in terms of true-positive fraction and false-positive fraction. Performance levels were compared among the acquisitions and reading modes. Time ...