Cervical spine: MR imaging with a partial flip angle, gradient-refocused pulse sequence. Part I. General considerations and disk disease.

1988 
A magnetic resonance imaging pulse sequence with a short repetition time (TR), short echo time (TE), partial flip angle, and gradient refocused echo was evaluated for the detection of cervical disk disease in a prospective study of 90 patients. These parameters were manipulated to adjust signal-to-noise ratio (S/N) and contrast: flip angle (3 degrees-18 degrees), TR (22-60 msec), and TE (12.5-25 msec). Flip angle had the greatest effect on S/N and contrast; its effect differed between axial and sagittal imaging. Cerebrospinal fluid S/N reached a peak at a smaller flip angle in sagittal imaging than in axial imaging. The useful range of flip angles depended on TR. Increasing TR had minimal direct effect on S/N or contrast, but because a longer TR allowed the use of larger flip angles for both axial and sagittal imaging, higher S/N could be achieved with similar contrast. This effect of increasing TR had to be balanced against increased imaging time and increased probability of motion artifact. Increasing T...