Reposition of a Bi-caval Dual Lumen ECMO cannula using a trans-femoral lasso

2019 
Abstract Veno-venous double lumen Extracorporeal membrane oxygenation (VVDL-ECMO) in neonates and infants for viral pneumonia is a well-established therapy. However even after initial correct positioning of the VVDL-ECMO cannula, there is a risk of unexpected malpositioning of this cannula during the ECMO run. We report a case of reposition of a 16-French Avalon Elite Bi-caval Dual Lumen cannula from its malposition in the right atrium to its correct position in the inferior venous cava. The procedure was achieved by interventional catheter techniques using a goose neck snare and was done without ECMO-flow interruption.
    • Correction