Evaluation of a rapid qualitative immuno-chromatography D-dimer assay (Simplify D-dimer) for the exclusion of pulmonary embolism in symptomatic outpatients with a low and intermediate pretest probability. Comparison with two automated quantitative assays.

2009 
Abstract The performance of a rapid qualitative solid-phase immuno-chromatography-based D-dimer assay (Simplify D-dimer) for diagnosing pulmonary embolism (PE) was evaluated in 469 outpatients with a low or intermediate pretest probability. They were referred to the emergency department of a university hospital during a 4-month period. Test results were compared to those of two automated quantitative assays. Simplify D-dimer assay result was positive in all 47 patients in whom the diagnosis of PE was retained and in 219 of the 422 patients without PE (51.2%), leading to a sensitivity of 100% (95%CI, 92.5 to 100%), a specificity of 48.8% (95%CI, 44.0 to 53.6%) and a negative predictive value (NPV) of 100% (95%CI, 98.2 to 100%). These results compared favorably with those of the Vidas D-dimer New assay [sensitivity = 100% (95%CI, 92.5 to 100), specificity = 49.8% (95%CI, 45.0 to 54.6%) and NPV = 100% (95%CI, 98.3 to 100%)] and the STA ® -Liatest ® D-DI assay [sensitivity = 100% (95%CI, 92.5 to 100%), specificity = 48.1% (95%CI, 43.3 to 52.9%) and NPV = 100% (95%CI, 98.2 to 100%)]. The inter-observer (n = 2) variability was very good with 1.7% discordant readings and a kappa coefficient ( K ) value = 0.97 (95%CI, 0.93 to 1.00). In conclusion, the Simplify D-dimer assay could be a valuable tool for ruling out PE in out-patients but a specific learning course of those having to work with is required in order to minimize the number of ambiguous reading and to overcome the inter-observer variability.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    40
    References
    6
    Citations
    NaN
    KQI
    []