Importance Strategies to procure high-quality core-needle biopsy (CNB) specimens are critical for making basic tissue diagnoses and for ancillary testing. Objectives To investigate acquisition of fluorescence confocal microscopy (FCM) images of interventional radiology (IR)–guided CNB in real time in the radiology suite and to compare the accuracy of FCM diagnoses with those of hematoxylin-eosin (HE these diagnoses were compared with those made using HE mean [SD] age, 63 [13] years) underwent IR-guided CNBs in a mean (SD) of 7 (2) minutes each. The FCM images showed at least 20% of the tissue with optimal quality in 101 CNB specimens (96.2%). The FCM images were accurately interpreted by the 2 pathologists in 100 of 105 cases (95.2%) (2 false-positive and 3 false-negative) and 90 of 105 cases (85.7%) (6 false-positive and 9 false-negative). A reassessment of 14 discordant diagnoses resulted in consensus diagnoses that were accurate in 101 of 105 cases (96.2%) (1 false-positive and 3 false-negative). Conclusions and Relevance The ease of acquisition of FCM images of acceptable quality and the high accuracy of the diagnoses suggest that FCM may be useful for rapid evaluation of IR-guided CNBs. This approach warrants further investigation.