Efficacy and validity of automated quantitative chemiluminescent enzyme immunoassay for SARS-CoV-2 antigen test from Saliva specimen in the diagnosis of COVID-19
2021
Abstract Introduction The pandemic of a novel coronavirus disease 2019 (COVID-19) caused by a severe acute respiratory coronavirus 2 (SARS-CoV-2) infection has been problematic worldwide. A new SARS-CoV-2 antigen test (LUMIPULSEⓇ) was licensed and widely used in Japan since May 2020. We conducted this study intending to whether the automated quantitative CLEIA antigen test using a saliva sample is effective and valid for the diagnosis of COVID-19. Patients and methods We analyzed and compared the diagnostic accuracy of both the automated quantitative CLEIA antigen test and real-time RT-PCR (rRT-PCR) using a saliva sample from individuals suspected as having COVID-19. Results A total of 305 samples were collected and tested in Aichi Medical University Hospital and affiliated facilities from December 2020 until January 2021 at our institute. Using reverse-transcription PCR as a reference, the AUROC of the automated quantitative CLEIA antigen test was 0.903 (95% confidential interval 0.845-0.962, p > 27, which shows the AUROC was 0.795 (95%CI 0.687-0.907, p Conclusion While the automated quantitative CLEIA antigen test from saliva specimen could be one of the most useful diagnostic tests for the diagnosis of COVID-19 in general practice, clinicians should know the limitations of the antigen test.