Three-year clinical outcomes with the ABSORB bioresorbable vascular scaffold in real life: Insights from the France ABSORB registry.

2020 
OBJECTIVES The aim of this study was to determine the 3-year outcomes of patients treated with Absorb bioresorbable vascular scaffold (BVS) implantation. BACKGROUND Randomized trials and observational registries performed in patients undergoing percutaneous coronary intervention have demonstrated higher 1-year and midterm rates of device thrombosis and adverse events with BVS compared to contemporary drug eluting stent. Data on long-term follow-up of patients treated with BVS are scarce. METHODS All patients treated with BVS were included in a large nationwide prospective multicenter registry (FRANCE ABSORB). The primary endpoint was a composite of cardiovascular death, myocardial infarction, and target lesion revascularization at 3 years. Secondary endpoints were 3-year scaffold thrombosis and target vessel revascularization (TVR). RESULTS Between September 2014 and April 2016, 2070 patients were included (mean age 55 ± 11 years; 80% men). The indication was acute coronary syndrome (ACS) in 49% of patients. At 3 years, the primary endpoint occurred in 184 patients (8.9%) and 3-year mortality in 43 patients (2.1%). Scaffold thrombosis and TVR rates through 3 years were, respectively, 3 and 7.6%. In a multivariable analysis, independent predictors of primary endpoint occurrence were diabetes, oral anticoagulation, active smoking, absence of initial angiographic success and the association of a total BVS length ≥30 mm with the use of 2.5 mm diameter BVS. CONCLUSIONS Although 3-year mortality was low in this ACS population, device-related events were significant beyond 1 year. Total BVS length and 2.5 mm BVS were associated with higher rates of MACE at long-term follow-up.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    27
    References
    1
    Citations
    NaN
    KQI
    []