Feasibility study of the Fricke chemical dosimeter as an independent dosimetric system for the small animal radiation research platform (SARRP).

2020 
Abstract For the small animal radiation research platform (SARRP) with X-ray beams in the medium energy range (tube operating voltage at 220 kVp), reference dosimetry is based on the AAPM TG-61 recommendations following the in-phantom method. The objective of this study was to evaluate the feasibility of the Fricke solution as a dosimeter to determine the absorbed dose to water. Feasibility studies at this X-ray energy range are not widely available. We evaluated the accuracy, dose linearity and dose rate dependence in a comparison with an NE 2571 Farmer ionization chamber (IC) and measurements in water. The G(Fe3+) factor was calculated from the curve fitting of the chemical yields for two radioactive sources (192Ir and 60Co) and one X-ray system with a tube operating at 150 and 250 kVp. The same methodology was followed for the dependence of the G(Fe3+) value on the energy and the dose agreement assessment for 180 and 200 kVp in the SARRP. The Fricke system exhibits a good linear response over the range of 5–70 Gy and an accuracy better than 2% for a 2 Gy/min dose rate. The dose rate dependence is smaller than 1% for dose rates greater than 1 Gy/min. The dependence of the G(Fe3+) value on the energy is smaller than 0.41%, with dose agreements better than 2%. The feasibility of the dosimeter for measurements at high doses and high dose rates makes it a suitable tool for dosimetric verifications in several preclinical irradiation configurations.