Passive-fixation lead failure rates and long-term patient mortality in subjects implanted with Sprint Fidelis electrodes

2014 
Aims To evaluate passive-fixation lead failure rates and long-term patient survival in subjects implanted with Sprint Fidelis electrodes. Methods and results We identified 748 subjects who received a Sprint Fidelis ( n = 429; Medtronic models 6948: 94.8%, 6949: 2.6%, 6930: 1.9%, 6931: 0.7%) or a Sprint ‘non-Fidelisimplantable cardioverter defibrillator lead ( n = 319, Medtronic models 6944: 68.6%, 6947: 17.9%, 6942: 7.8%, 6943: 3.4%, 6945: 2.2%) at our centre between 1998 and 2008. Kaplan–Meier patient survival was lower in the Fidelis group than in the Control cohort (68.4 vs. 77.0% at 5 years, P = 0.0061), but multivariate analyses revealed no significant association between mortality and implanted lead type. Passive-fixation lead failure rate at 5 years was 14.4% (95% confidence interval (CI) [9.2, 19.3]) in the Fidelis ( n = 414) group and 1.8% (95% CI [0.0–3.8]) in the Control ( n = 241) cohort ( P < 0.001 upon multivariate comparison). Conclusion Failure rates of passive-fixation Sprint Fidelis leads are increased and similar to those previously reported for active-fixation Fidelis electrodes. Despite the elevated risk for lead failure and its potential sequelae, the Sprint Fidelis has no obvious impact on long-term mortality.