Development of a sensitive induction-based magnetic nanoparticle biodetection method

2018 
We developed a novel biodetection method for influenza virus based on AC magnetic susceptibility measurement techniques (the DynoMag induction technique) together with functionalized multi-core magnetic nanoparticles. The sample consisting of an incubated mixture of magnetic nanoparticles and rolling circle amplified DNA coils is injected into a tube by a peristaltic pump. The sample is moved as a plug to the two well-balanced detection coils and the dynamic magnetic moment in each position is read over a range of excitation frequencies. The time for making a complete frequency sweep over the relaxation peak is about 5 minutes (10 Hz–10 kHz with 20 data points). The obtained standard deviation of the magnetic signal at the relaxation frequency (around 100 Hz) is equal to about 10−5 (volume susceptibility SI units), which is in the same range obtained with the DynoMag system. The limit of detection with this method is found to be in the range of 1 pM.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    10
    References
    5
    Citations
    NaN
    KQI
    []