Can voltammetry measure nitrogen monoxide (NO) and/or nitrites?

2001 
Abstract Recently, voltammetry with carbon fibre electrodes (CFE) has been implemented for real time measurement of nitrogen monoxide (NO) indicating that it is oxidised at the potential value of nitrites, approximately +700 mV. In contrast, here we show that modified CFE can monitor NO at oxidation potentials different than that of nitrites, i.e. +550 mV. Indeed, at +550 mV a significant increase of amperometric current levels was obtained when NO but not nitrites, were added to a phosphate buffer saline solution (PBS). Differential pulse voltammetry (DPV) supports these findings as two oxidation peaks were obtained when examining air preserved NO; peak 1 at +550 mV and peak 2 at +700 mV, respectively. In contrast, only peak 2 was monitored when nitrites or a solution of NO oxidised in air was added to PBS. Biological support to these in vitro data comes from the observation that the relaxation of an adrenaline-contracted aortic ring produced via addition of NO is concomitant with peak 1 at +550 mV. The relaxation is almost completed before the appearance of peak 2 at +700 mV. Furthermore, in vivo experiments performed in the striatum of rats show that the amperometric signal monitored at +550 mV is responsive to glutamatergic stimulation or inhibition of NO synthase.