Thin co-radial bipolar leads: technology and clinical performance
2001
While bipolar leads offer advantages such as better sensing performance than unipolar leads, their use has been limited by a larger lead diameter and reports about a high failure rate of several bipolar lead models. This has led to the development of thin bipolar leads using a special technology which aims at improving lead safety. Leads with monofilar thin conductors (drawn filled tube) which are individually coated with a very resistant material (ETFE) have been developed. Using a co-radial instead of co-axial bipolar conductor design, the lead diameter could be reduced to 4.5 F compared to 6–7 F of conventional bipolar leads. Bench testing demonstrated a significant improvement of this lead technology with respect to degradation of insulation material by biochemically reactive solutions. Also mechanical characteristics such as resistance to tearing forces and compression showed a high lead durability. From our own experience, co-radial bipolar leads show a favorable electrical performance with the exception of a relatively low pacing impedance. Also during long-term follow-up, the rate of lead failure was very low. These findings are corroborated by other clinical studies which also demonstrated good handling characteristics of thin bipolar leads during implantation.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
22
References
3
Citations
NaN
KQI