Glutaraldehyde-Fixed Bioprosthetic Heart Valve Conduits Calcify and Fail From Xenograft Rejection
2006
Background— Glutaraldehyde fixation (G-F) decreases but likely does not eliminate the antigenicity of bioprosthetic heart valves. Rejection (with secondary dystrophic calcification) may be why G-F xenograft valves fail, especially in young patients, who are more immunocompetent than the elderly. Therefore, we sought to determine whether rejection of G-F xenograft occurs and to correlate this with graft calcification. Methods and Results— Ascending aortas/valves (from rats [syngeneic] or guinea pigs [xenogeneic]) were transplanted (fresh or after 48 hour of G-F) into the infrarenal aortas of young rat recipients for 20 days. A xenogeneic group was also treated with steroids until graft harvest. The valves and media/adventitia were scored blindly for inflammation (0 to 4). Percent graft infiltration by T cells/macrophages was determined (immunohistochemistry), and rat IgG ELISAs were performed. There was >3 times more valve inflammation, >10 times more valve T-cell/macrophage infiltrate, and >3 times antibo...