Temperature monitoring during radiofrequency catheter ablation procedures using closed loop control. Atakr Multicenter Investigators Group.

1994 
BACKGROUNDThe purpose of this study was to evaluate electrode temperatures obtained using a radiofrequency ablation system that incorporates closed loop feedback control to achieve preset target electrode temperatures and to determine if closed loop temperature control results in a lower incidence of developing a coagulum.METHODS AND RESULTSTwo hundred seventy patients underwent catheter ablation of atrioventricular nodal reentrant tachycardia, an accessory pathway, and/or the atrioventricular junction using an ablation system incorporating closed loop feedback control. Forty-five patients underwent catheter ablation in the power control mode in which power output was fixed, and 225 patients underwent catheter ablation in the temperature control mode. A coagulum occurred during 0.8% of radiofrequency applications in the temperature control mode versus 2.2% in the power control mode (P < .01). Electrode temperatures were within 10 degrees C of the targeted temperature during 35% of applications in the temp...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    11
    References
    142
    Citations
    NaN
    KQI
    []