Effect of bipolar electrode spacing on phrenic nerve stimulation and left ventricular pacing thresholds: an acute canine study.

2012 
Background—Phrenic nerve stimulation (PNS) is a common complication of cardiac resynchronization therapy when left ventricular (LV) pacing occurs via a coronary vein. The purpose of this study was to evaluate the effects of bipolar electrode spacing on PNS and LV pacing thresholds. Methods and Results—Electrophysiology catheters with standard (2 mm-5 mm-2 mm) or modified (1 mm-5 mm-1 mm) interelectrode spacing was, respectively, inserted in a posterior/lateral cardiac vein in a randomized order in 6 anesthetized dogs via jugular access. The phrenic nerve was dissected via a left minithoracotomy and repositioned over the vein as close as possible to one of the electrodes. The presence of PNS was verified (ie, PNS threshold 0.05). Compared with a standard bipolar electrode spacing of 20 mm for LV leads, 1 and 2 mm bipolar electrode spacing resulted in a PNS threshold increase of 5.5±2.2 V (P=0.003) and 2.8±1.7 V (P<0.001), respectively. Similarly, PNS threshold increased by 6.5±3.7 V with 1 mm and by 3.8±1.9 V with 2 mm bipolar pacing (both P<0.001), compared with unipolar pacing. Conclusions—This study suggests that reducing LV bipolar electrode spacing from the standard 20 mm to 1 or 2 mm may significantly increase the PNS threshold without compromising LV pacing thresholds. (Circ Arrhythm Electrophysiol. 2012;5:815-820.)