Diagnostic Ability of Automated Pupillography in Glaucoma

2017 
ABSTRACTPurpose: To evaluate the diagnostic ability of automated pupillography measurements in glaucoma and study the effect of inter-eye asymmetry in glaucomatous damage on the diagnostic ability.Methods: In an observational, cross-sectional study, 47 glaucoma patients and 42 control subjects underwent automated pupillography using a commercially available device. Diagnostic abilities of the pupillary response measurements were evaluated using area under receiver operating characteristic (ROC) curves (AUC) and sensitivities at fixed specificities. Influence of inter-eye asymmetry in glaucoma [inter-eye mean deviation (MD) difference on visual fields (VF)] on the diagnostic ability of pupillography parameters was evaluated by ROC regression approach.Results: The AUCs of automated pupillography parameters ranged from 0.60 (amplitude score with peripheral blue stimulus) to 0.82 (amplitude score with full field white stimulus, Amp-FF-W). Sensitivity at 95% specificity ranged between 5% (amplitude score with ...