Late Adverse Events After Implantation of Sirolimus-Eluting Stent and Bare-Metal Stent Long-Term (5–7 Years) Follow-Up of the Coronary Revascularization Demonstrating Outcome Study-Kyoto Registry Cohort-2

2014 
Background—Late adverse events such as very late stent thrombosis (VLST) or late target-lesion revascularization (TLR) after first-generation sirolimus-eluting stents (SES) implantation have not been yet fully characterized at long term in comparison with those after bare-metal stent (BMS) implantation. Methods and Results—Among 13 058 consecutive patients undergoing first percutaneous coronary intervention in the Coronary REvascularization Demonstrating Outcome study-Kyoto registry Cohort-2, 5078 patients were treated with SES only, and 5392 patients were treated with BMS only. During 7-year follow-up, VLST and late TLR beyond 1 year after SES implantation occurred constantly and without attenuation at 0.24% per year and at 2.0% per year, respectively. Cumulative 7-year incidence of VLST was significantly higher in the SES group than that in the BMS group (1.43% versus 0.68%, P<0.0001). However, there was no excess of all-cause death beyond 1 year in the SES group as compared with that in the BMS group (...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    33
    References
    64
    Citations
    NaN
    KQI
    []