Design and optimizing a novel ocular plaque brachytherapy with dual-core of 103Pd and 106Ru.

2021 
Abstract In recent decades, eye plaques of brachytherapy have been extensively used as primary treatment as well as a complementary treatment for ocular cancer. The purpose of this study is the development of the eye plaque brachytherapy throughout a new design of eye plaque by combining the COMS plaque and the CCB BEBIG plaque loaded by IRA1-103Pd and 106Ru, respectively. A new dual-core plaque with a diameter of 20 mm was designed in the way that the BEBIG plaque with a diameter of 20 mm loaded by 106Ru plate is attached to the COMS plaque with a diameter of 20 mm loaded by 24 of IRA1-103Pd seeds. Dose calculations for the new plaque were performed by using the MCNP5 code. Dose calculations of dual-core plaque including 103Pd seeds (gamma) and 106Ru plate (beta) were separately done for the sake of MCNP constraints in gamma and beta particle transfer simultaneously. The new dual-core plaque delivers a much higher dose rate to the tumor compared with every single plaque, while the dose rate reached to healthy tissues is slightly higher than each plaque separately. Of course, this is acceptable because the treatment time reduces and subsequently the error in radiation therapy reduces.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    60
    References
    0
    Citations
    NaN
    KQI
    []