Improved mitral valve coaptation and reduced mitral valve annular size after percutaneous mitral valve repair (PMVR) using the MitraClip system

2018 
Aims: Improved mitral valve leaflet coaptation with consecutive reduction of mitral regurgitation (MR) is a central goal of percutaneous mitral valve repair (PMVR) with the MitraClip® system. As influences of PMVR on mitral valve geometry have been suggested before, we examined the effect of the procedure on mitral annular size in relation to procedural outcome. Methods and results: Geometry of the mitral valve annulus was evaluated in 183 patients undergoing PMVR using echocardiography before and after the procedure and at follow-up. Mitral valve annular anterior-posterior (ap) diameter decreased from 34.0 ± 4.3 to 31.3 ± 4.9 mm (P < 0.001), and medio-lateral (ml) diameter from 33.2 ± 4.8 to 32.4 ± 4.9 mm (P < 0.001). Accordingly, we observed an increase in MV leaflet coaptation after PMVR. The reduction of mitral valve ap diameter showed a significant inverse correlation with residual MR. Importantly, the reduction of mitral valve ap diameter persisted at follow-up (31.3 ± 4.9 mm post PMVR, 28.4 ± 5.3 mm at follow-up). Conclusion: This study demonstrates mechanical approximation of both mitral valve annulus edges with improved mitral valve annular coaptation by PMVR using the MitraClip® system, which correlates with residual MR in patients with MR.