Nanoduct ® sweat testing for rapid diagnosis in newborns, infants and children with cystic fibrosis

2008 
Determination of chloride concentration in sweat is the current diagnostic gold standard for Cystic Fibrosis (CF). Nanoduct® is a new analyzing system measuring conductivity which requires only 3 microliters of sweat and gives results within 30 minutes. The aim of the study was to evaluate the applicability of this system in a clinical setting of three children’s hospitals and borderline results were compared with sweat chloride concentration. Over 3 years, 1,041 subjects were tested and in 946 diagnostic results were obtained. In 95 children, Nanoduct® failed (9.1% failure rate), mainly due to failures in preterm babies and newborns. Assuming 59 mmol/L as an upper limit of normal conductivity, all our 46 CF patients were correctly diagnosed (sensitivity 100%, 95% CI: 93.1–100; negative predicted value 100% (95% CI: 99.6–100) and only 39 non CF’s were false positive (39/900, 4.3%; specificity 95.7%, 95%CI: 94.2–96.9, positive predicted value 54.1% with a 95%CI: 43.4–65.0). Increasing the diagnostic limit to 80 mmol/L, the rate fell to 0.3% (3/900). CF patients had a median conductivity of 115 mmol/L; the non-CF a median of 37 mmol/L. In conclusion, the Nanoduct® test is a reliable diagnostic tool for CF diagnosis: It has a failure rate comparable to other sweat tests and can be used as a simple bedside test for fast and reliable exclusion, diagnosis or suspicion of CF. In cases with borderline conductivity (60–80 mmol/L) other additional methods (determination of chloride and genotyping) are indicated.