Dosimetric comparison of helical tomotherapy, volumetric-modulated arc therapy, intensity-modulated radiotherapy, and field-in-field technique for synchronous bilateral breast cancer

2020 
ABSTRACT Purpose: To compare the dosimetric characteristics of helical tomotherapy (HT), volumetric-modulated arc therapy (VMAT), intensity-modulated radiotherapy (IMRT), and tangential field-in-field technique (FIF) for the treatment of synchronous bilateral breast cancer (SBBC). Methods and Materials: Ten patients with early-stage unilateral breast cancer were selected for simulating the patients with SBBC in this retrospective analysis. Treatment plans with HT, VMAT, IMRT, and FIF were generated for each patient with a total dose of 50.4 Gy in 28 fractions to the target. Plan quality, namely conformity index (CI), homogeneity index (HI), dose-volume statistics of organs at risk (OARs), and beam-on time (BOT), were evaluated. Results: HT plans showed a lower mean heart dose (3.53 ± 0.31Gy) compared with the other plans (VMAT = 5.6 ± 1.36 Gy, IMRT = 3.80 ± 0.76 Gy, and FIF = 4.84 ± 2.13 Gy). Moreover, HT plans showed a significantly lower mean lung dose (p