COMPARATIVE RETROSPECTIVE STUDY OF 320 PRIMARY CHARNLEY TOTAL HIP ARTHROPLASTY WITH A MINIMUM OF TEN YEARS' FOLLOW-UP TO ASSESS WHETHER DUAL MOBILITY CUP HAS A DECREASED DISLOCATION

2018 
Mid and long-term follow-up of Charnley total hip arthroplasty (THA) demonstrated good functional results with 85% survivorship at 25-year follow-up. However dislocation still remains an unsolved problem. Dislocation may occur all along the patient and implant life. The aim of this study is to answer the question: does Dual Mobility Cup (DMC) decrease the dislocation risk? Method: We report comparative results at ten years of follow-up of 2 groups of primary cemented Charnley-type THA, one with a standard polyethylene cup (group 1, n=215) and the other one with a DMC (group 2, n=105). Results: In group 1, twenty-six dislocations (12.9%) occurred. In group 2 only one dislocation (0.9%) occurred. This dislocation was successfully reduced by close reduction, without any recurrence. This difference was statistically significant (p=0.0018). In group 1, reason for revision was recurrent dislocation in twenty one cases. Five patients have been revised for other reasons. The global revision rate was 12.9%. In gro...