Platelet activation is less enhanced in the new balloon expandable Edwards Sapien 3 valve compared to its predecessor model (Edwards Sapien XT)

2015 
Stroke and thromboembolic events after transfemoral aortic valve replacement (TAVR) continue to be a problem. The aim of our study was to compare platelet aggregation (Agg) and platelet activation (PA) observed with two different catheter valves, the ESV-XT and the newer ESV-3 valve in patients (pts) undergoing TAVR on dual antiplatelet therapy (DAPT). A total of 174 patients with severe aortic stenosis and high surgical risk successfully underwent TAVR (60 ESV-XT; 114 ESV-3). Platelet Agg and PA (CD62P expression) were evaluated before and the following three days after TAVR under DAPT. Platelet Agg was inhibited to the same extent in both valve types and there was no significant difference in platelet drop between both valve types between day 0 and day 3 [ESV-XT vs ESV-3: median (25th-75th percentile): platelet count (x1000): 55 (42–74) vs 61(42–93), p=0.280]. However, there was an enhanced CD62P expression directly after TAVR with the ESV-XT compared to the ESV-3 [CD62P (MIF): 7.4 (6.8–8.6) vs 6.6 (6–7.9), p=0.014]. Surface expression of platelet CD62P was associated with the occurrence of residual aortic regurgitation (AR) and was significantly higher in patients with residual AR [CD62P (mild AR) vs CD 62P (no or trace AR): 7.9 (7.3–9.1) vs 7.1 (6.4–8.0), p
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    7
    Citations
    NaN
    KQI
    []