Correlation between the VITEK2 system and cefoxitin disk diffusion for the daily detection of oxacillin resistance in a large number of clinical Staphylococcus aureus isolates

2010 
The aim of the present study was to compare the performance of the new VITEK2 AST-P551 card with the cefoxitin disk diffusion method for the daily detection of methicillin resistance with a high number of Staphylococcus aureus clinical isolates. Detection of the PBP2a protein or mecA gene was performed for each discordant case. Seventy (3.3%) isolates out of 2,107 clinical strains showed discordant results, two very major errors, four major errors and 64 minor errors. Fifty-nine (84%) discordant results were resolved, with a final overall agreement of 99.5%. Eleven (0.5%) strains remained discordant (minor error [mE]). Four of 370 MRSA strains were misclassified as susceptible in daily practice by the cefoxitin disk diffusion method. All of these strains were resistant to aminoglycosides and/or fluoroquinolones. The VITEK2 system is highly reliable for methicillin resistance detection at the routine level. Oxacillin-susceptible classified clinical strains with associated resistance patterns required attention.