Non-invasive electrophysiological assessment of the optimal configuration of quadripolar lead vectors on ventricular activation times

2018 
Abstract Background Cardiac resynchronization therapy (CRT) is now generally delivered via quadripolar leads. Assessment of the effect of different vector programs from quadripolar leads on ventricular activation can be now done using non-invasive electrocardiographic mapping (ECM). Material and methods In nineteen patients with quadripolar LV leads, activation maps were constructed. The total ventricular activation time (TVaT) and the time for the bulk of ventricular activation (VaT 10–90 ) were calculated. Results CRT delivered via a quadripolar lead significantly reduced TVaT and VaT 10–90 by a mean of 16 ms and 31 ms, respectively, compared to baseline. There was a marked reduction in ventricular activation between the most and least synchronous vectors: 28% difference in baseline TVaT and 37% difference in VaT 10–90 . Conclusion Changes in the configuration of an LV quadripolar lead significantly affected ventricular activation timings in both ischaemic and non-ischaemic subjects. This suggests that programming of the optimal pacing vector may need to be individually tailored.