Experimental Investigation of Aortic Pressure Variations Following a Simulated Thoracic Impact.

2021 
Blunt traumatic aortic rupture is a heart injury that can occur in falls, automobile accidents, and sporting injuries involving impact to the thorax. Despite its severity and high morbidity rate, the research still does not provide a consistent description of the mechanism of rupture. In this study, a crash testing dummy with an in vitro pumping heart, 3D printed ribcage, and ballistic gel damping layer was developed to reproduce a realistic response to thoracic impact. Testing was performed using a standardized pendulum used for calibration of crash test dummies, with the location of impact being the middle of the sternum. Different impact severities were tested by adjusting the kinetic energy at impact with the initial height of the pendulum. Measurements of the dummy include instantaneous aortic pressure waveforms prior, during and following the impact. The results of this experiment show that aortic pressure experiences significant changes in magnitude during simulated impact. This work could help contribute towards a better understanding of the mechanisms leading to blunt traumatic aortic rupture and the development of preventative measures.