A Sutureless Vascular Closure Device for Emergent Bovine Xenograft Implantation

2017 
ABSTRACTA novel vascular staple (C-staple) was developed that does not enter the vasculature lumen during anastomoses. The objective of this study was to demonstrate C-staple safety when used with a bovine xenograft and compare efficacy of the C-staple procedure with Anastoclip surgical clips or suturing when used with a bovine xenograft. Eight sheep had an acute comparison between suturing and C-staples using both common carotid arteries. Sixteen sheep had xenograft placement in the left carotid artery, eight with C-staples and eight with Anastoclips in a chronic study. Over 6 months, Doppler ultrasound interrogation of the common carotid arteries was performed. After 6 months, arteries were evaluated histopathologically. Cross-clamp and surgical times were longer in the C-staple group than the suture group, and xenograft implantation times were statistically longer with C-staples than with Anastoclips. After 6 months, C-staple biocompatibility was similar to Anastoclips. Patency and hemodynamics of the ...