2013 
Objectives:To estimate the risk of injury from radiofrequency (RF) heating of metallic dental devices in use during 3.0 T MRI.Methods:The whole-body specific absorption rate (WB-SAR) was calculated on the basis of saline temperature elevation under the maximum RF irradiation for 15 min to determine the operation parameters for the heating test. The temperature changes of three types of three-unit bridges, a full-arch fixed dental prosthesis and an orthodontic appliance in use during MRI with a 3.0 T MR system (Magnetom® Verio; Siemens AG, Erlangen, Germany) were then tested in accordance with the American Society for Testing and Materials F2182-09 standardized procedure under the maximum RF heating during 15 min RF irradiation.Results:The system console-predicted WB-SAR was approximately 1.4 W kg−1 and that measured with a saline phantom was 2.1 W kg−1. In the assessment of RF heating, the highest temperature increase was +1.80 °C in the bridges, +1.59 °C in the full-arch fixed dental prosthesis and +2.61...