Electrical-thermal performance of a cooled RF applicator for hepatic ablation with additional distant infusion of hypertonic saline: in vivo study and preliminary computer modeling.

2012 
Purpose: The Cool-tip electrode is one of the most widely employed applicators in radiofrequency (RF) hepatic ablation. Previous research demonstrated that it is possible to enlarge coagulation volume when the single cooled electrode is associated with distant infusion of saline (hybrid applicator). The aim of this study was to compare the electrical-thermal behaviour of the Cool-tip electrode with that of the hybrid applicator.Materials and methods: Forty-two RF ablations were performed on a total of 10 pigs: 22 with the Cool-tip electrode and 20 with the hybrid applicator (low infused saline volumetric flow rate of 6 mL/h at 2 mm distance). We compared both electrical performance (delivered power and number of roll-offs, i.e. sudden rises in impedance that interrupt the power delivery) and coagulation zone characteristics. In addition, we built a one-dimensional model to provide a basic physical explanation of the difference in performance between the different applicators.Results: The experimental resu...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    11
    Citations
    NaN
    KQI
    []