Analytical and Clinical Validation of a Point-of-Care Cardiac Troponin T Test with an Improved Detection Limit.

2017 
Abstract The point-of-care test Roche CARDIAC POC Troponin T (PoC TnT) is an improved assay which has been developed for the Roche cobas h 232 system. We performed a multicentre evaluation (four sites) to assess the analytical performance of the PoC TnT assay and to compare it with the central laboratory Elecsys® troponin T high sensitive (lab cTnT-hs) assay. The relative mean differences found in method comparisons of PoC TnT vs. lab cTnT-hs ranged from -4.1% to +6.8%. Additionally, there was good concordance between PoC TnT and lab cTnT-hs for the number of samples with troponin T values below the measuring range of 40 ng/L. Lot-to-lot differences of PoC TnT ranged from -8.6% to +4.6%. Within-series coefficients of variation (CV) resulting from 81 ten-fold measurements with patient samples were 9.3%, 11.8%, and 12.9% in the low (40 to