Evolution of diagnostic reference levels in Spanish intraoral radiology.

2012 
: A total of 16 175 official reports of quality assurance on dental radiodiagnostic surgeries from 16 Spanish autonomous regions compiled during 2002-09 were studied to determine the evolution of diagnostic reference levels (DRLs) for obtaining a diagnostic image in the normal conditions of clinical practice in Spanish dental clinics. A DRL of 3.1 mGy was set in 2009, which represents a 35.4 % decrease compared with the dose determined in 2002 (4.8 mGy). During the same period, the mean dose fell by only 17.2 %. The DRL recommended by the European Union in 2004 for intraoral radiology is 4 mGy, and this study shows that 83.4 % of the installations used a dose below this. Of the installations using indirect or direct digital systems 1.1 and 1.2 %, respectively, used doses higher than those recommended, while 14.2 % of those using radiographic film exceeded this limit.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17
    References
    3
    Citations
    NaN
    KQI
    []