Automated office blood pressure measurements in primary care are misleading in more than one third of treated hypertensives: The VALENTINE-Greece Home Blood Pressure Monitoring study

2019 
Abstract Background This study assessed the diagnostic reliability of automated office blood pressure (OBP) measurements in treated hypertensive patients in primary care by evaluating the prevalence of white coat hypertension (WCH) and masked uncontrolled hypertension (MUCH) phenomena. Methods Primary care physicians, nationwide in Greece, assessed consecutive hypertensive patients on stable treatment using OBP (1 visit, triplicate measurements) and home blood pressure (HBP) measurements (7 days, duplicate morning and evening measurements). All measurements were performed using validated automated devices with bluetooth capacity (Omron M7 Intelli-IT). Uncontrolled OBP was defined as ≥140/90 mmHg, and uncontrolled HBP was defined as ≥135/85 mmHg. Results A total of 790 patients recruited by 135 doctors were analyzed (age: 64.5 ± 14.4 years, diabetics: 21.4%, smokers: 20.6%, and average number of antihypertensive drugs: 1.6 ± 0.8). OBP (137.5 ± 9.4/84.3 ± 7.7 mmHg, systolic/diastolic) was higher than HBP (130.6 ± 11.2/79.9 ± 8 mmHg; difference 6.9 ± 11.6/4.4 ± 7.6 mmHg, p  Conclusions In primary care, automated OBP measurements are misleading in approximately 40% of treated hypertensive patients. HBP monitoring is mandatory to avoid overtreatment of subjects with WCH phenomenon and prevent undertreatment and subsequent excess cardiovascular disease in MUCH.