Efficacy and Safety Results with the LifeSite Hemodialysis Access System versus the Tesio-Cath Hemodialysis Catheter at 12 Months

2006 
PURPOSE To compare the performance and safety of a fully subcutaneous vascular access device, the LifeSite hemodialysis access system, versus a tunneled hemodialysis catheter, the Tesio-Cath, at 1 year after implantation MATERIALS AND METHODS Sixty-eight patients who required hemodialysis received implantation of the LifeSite device or a Tesio-Cath device as a part of this multicenter study. Thirty-four patients were treated in each group. The endpoints observed included blood flow rates and associated venous pressures, overall and device-related adverse events, the need for thrombolytic infusions, device-related infections (DRIs) and associated hospitalizations, and technical device survival RESULTS During the 12-month observation period, significantly higher venous pressures were required in patients with the Tesio-Cath to achieve blood flow rates comparable with those achieved with the LifeSite device. Patients in the LifeSite group experienced a significantly lower rate of non–device-related adverse events ( P P P P P CONCLUSIONS The results of the present study demonstrate superior device performance and technical device survival, reduced complications, and the need for fewer interventions with the LifeSite hemodialysis access system compared with a standard hemodialysis catheter during a 1-year time period after implantation