Biomechanical behavior of titanium and zirconia frameworks for implant‐supported full‐arch fixed dental prosthesis
2017
Background
The biomechanical behavior of implant-supported titanium and zirconia full-arch fixed dental prosthesis (FAFDP) frameworks require further investigation.
Purpose
Strains transferred by implant-supported titanium (Ti) and zirconia (Zr) FAFDP frameworks were analyzed.
Materials and Methods
Maxillary 14-unit FAFDPs supported by 6 implants and 12-unit FAFDPs supported by 4 implants were tested. One-piece frameworks were fabricated by computer-aided design/computer-aided manufacturing. Four groups were divided (n = 3): G1, Ti-6 implants; G2, Zr-6 implants; G3, Ti-4 implants; G4, Zr-4 implants. A 250 N single-point load was applied on the second premolar. A three-dimensional digital image correlation system recorded framework and maxilla model surface deformation.
Results
The following strains (μS) averaged over the length of the second premolar were calculated: frameworks, G1 (321.82 ± 111.29), G2 (638.87 ± 108.64), G3 (377.77 ± 28.64), G4 (434.18 ± 132.21); model surface, G1 (473.99 ± 48.69), G2 (653.93 ± 45.26), G3 (1082.50 ± 71.14), G4 (1218.26 ± 230.37). Zirconia frameworks supported by 6 implants (G2) presented higher surface strains (P .05).
Conclusions
Zirconia frameworks supported by 6 implants showed higher strains. FAFDPs supported by 6 implants transferred less strains to the supporting maxilla, irrespective of framework material.