A prolonged observational study of tracheal tube displacements: Benchmarking an incidence <0.5-1% in a medical-surgical adult intensive care unit.

2014 
Background and Aims: Tracheal tubes are commonly used in intensive care unit (ICU) and lead to complications like displacements. The primary aim of the study was to evaluate if the rate of tracheal tube displacement benchmarked at Subjects and Methods: This was a prospective observational study of Intubated and ventilated patients in a General Medical-Surgical Adult ICU. The incidence of accidental extubation, self extubation, partial displacement and blockages of tracheal tubes were recorded. Results: The overall tracheal tube displacement rate was 61/10,112 (0.6%) per patient and 61/28,464 (0.22%) per tracheal tube day. There were 30 additional incidents of blockage, kinking or biting of the tracheal tube. Physiological consequences-69 were mild, 10 moderate, 12 major and one death. Of the 91 accidents, 30 were partly and 30 were completely preventable. 76 incidents involved an endotracheal tube (54 displaced, 12 blocked and 10 bitten-kinked) and 15 a tracheostomy tube (seven displaced and eight blocked). Accidents were more common in medical than surgical patients (medical = 48, cardiac surgical = 17 and other surgical/trauma = 26). Conclusion: Tracheal tube displacement rate in a mixed medical-surgical adult ICU was maintained below the pre-set benchmark of <1% per patient and <0.5% per intubated day over nearly a decade.