A pilot experience launching a national dose protocol for vascular and interventional radiology.

2008 
The design of a national dose protocol for interventional radiology has been one of the tasks during the European SENTINEL Coordination Action. The present paper describes the pilot experience carried out in cooperation with the Spanish Society on Vascular and Interventional Radiology (SERVEI). A prospective sample of procedures was initially agreed. A common quality control of the X-ray systems was carried out, including calibration of the air kerma area product (KAP) meters. Occupational doses of the radiologists involved in the survey were also included in the survey. A total of 10 Spanish hospitals with interventional X-ray units were involved. Six hundred and sixty-four patient dose data were collected from 397 diagnostic and 267 therapeutic procedures. Occupational doses were evaluated in a sample of 635 values. The obtained KAP median/mean values (Gy·cm 2 ) for the gathered procedures were: biliary drainage (30.6/68.9), fistulography (4.5/9.8), lower limb arteriography (52.2/60.7), hepatic chemoembolisation (175.8/218.3), iliac stent (45.9/73.2) and renal arteriography (39.1/59.8). Occupational doses (mean monthly values, in mSv) were 1.9 (over apron); 0.3 (under apron) and 4.5 (on hands). With this National experience, a protocol was agreed among the SENTINEL partners to conduct future similar surveys in other European countries.