Improvement of dose homogeneity with irregular surface compensator in whole brain radiotherapy

2016 
Purpose The aim of this study was to evaluate the dosimetric aspects of whole brain radiotherapy (WBRT) using an irregular surface compensator (ISC) in contrast to conventional radiotherapy techniques. Methods Treatment plans were devised for 20 patients. The Eclipse treatment planning system (Varian Medical Systems) was used for dose calculation. For the ISC, a fluence editor application was used to extend the range of optimal fluence. The treatment plan with the ISC was compared with the conventional technique in terms of doses in the planning target volume (PTV), dose homogeneity index (DHI), three-dimensional (3D) maximum dose, eye and lens doses and monitor unit (MU) counts required for treatment. Results Compared with conventional WBRT, the ISC significantly reduced the DHI, 3D maximum dose and volumes receiving 105% of the prescription dose, in addition to reducing both eye and lens doses ( p p Conclusion The ISC technique for WBRT considerably improved the dose homogeneity in the PTV. As patients who receive WBRT have improved survival, the long-term side effects of radiotherapy are highly important.