Reliability and Limitations of Automated Arrhythmia Detection in Telemetric Monitoring After Stroke
2015
Background and Purpose—Guidelines recommend continuous ECG monitoring in patients with cerebrovascular events. Studies on intensive care units (ICU) demonstrated high sensitivity but high rates of false alarms of monitoring systems resulting in desensitization of medical personnel potentially endangering patient safety. Data on patients with acute stroke are lacking. Methods—One-hundred fifty-one consecutive patients with acute cerebrovascular events were prospectively included. Automatically identified arrhythmia events were analyzed by manual ECG analysis. Muting of alarms was registered. Sensitivity was evaluated by beat-to-beat analysis of the entire recorded ECG data in a subset of patients. Ethics approval was obtained by University of Erlangen-Nuremberg. Results—A total of 4809.5 hours of ECG registration and 22 509 alarms were analyzed. The automated detection algorithm missed no events but the overall rate of false alarms was 27.4%. Only 0.6% of all alarms indicated acute life-threatening events ...
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
14
References
10
Citations
NaN
KQI