Test of UFSD Silicon Detectors for the TOTEM Upgrade Project

2017 
This paper describes the performance of a prototype timing detector, based on 50 micrometer thick Ultra Fast Silicon Detector, as measured in a beam test using a 180 GeV/c momentum pion beam. The dependence of the time precision on the pixel capacitance and the bias voltage is investigated here. A timing precision from 30 ps to 100 ps, depending on the pixel capacitance, has been measured at a bias voltage of 180 V. Timing precision has also been measured as a function of the bias voltage.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    11
    References
    1
    Citations
    NaN
    KQI
    []