Prevention of hypoxia and hyperoxia during endotracheal suctioning

1987 
: A new suction catheter, designed to deliver alternately oxygen or suction, prevented episodes of hypoxia and hyperoxia in a group of infants during endotracheal suctioning. Twenty infants received both conventional endotracheal suctioning and suctioning by the new catheter. The infants had a maximal change from a presuctioning transcutaneous oxygen (PtcO2) of 12 +/- 8 torr and required 3.1 +/- 2 min to regain their presuctioning oxygenation level compared to a maximal change of 21 +/- 10 torr (p less than .05) and a stabilization time of 5.3 +/- 2.6 min (p less than .05) in the conventionally treated group. Three study infants experienced an abnormal PtcO2 (either less than 40 or greater than 90 torr), while 13 control infants suffered these abnormalities (p less than .01). The use of this new suction device effectively reduced the exposure of this group of infants to episodes of aberrant oxygen states and allowed for a shorter recovery time.