Coplanar versus noncoplanar intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) treatment planning for fronto-temporal high-grade glioma†

2012 
The purpose of this study was to compare dosimetric and radiobiological parameters of treatment plans using coplanar and noncoplanar beam arrangements in patients with fronto-temporal high-grade glioma (HGG) generated for intensity-modulated radiotherapy (IMRT) or volumetric-modulated arc therapy (VMAT). Ten cases of HGG overlapping the optic apparatus were selected. Four separate plans were created for each case: coplanar IMRT, noncoplanar IMRT (ncIMRT), VMAT, and noncoplanar VMAT (ncVMAT). The prescription dose was 60 Gy in 30 fractions. Dose-volume histograms and equivalent uniform doses (EUD) for planning target volumes (PTVs) and organs at risk (OARs) were generated. The four techniques resulted in comparable mean, minimum, maximum PTV doses, and PTV EUDs ( p ≥ 0.33). The mean PTV dose and EUD averaged for all techniques were 59.98Gy (Standard Deviation (SD) ± 0.15) and 59.86 Gy (SD ± 0.27). Non­coplanar IMRT significantly reduced contralateral anterior globe EUDs (6.7 Gy versus 8.2 Gy, p = 0.05), while both ncIMRT and ncVMAT reduced contralateral retina EUDs (16Gy versus 18.8 Gy, p = 0.03). Noncoplanar techniques resulted in lower contralateral temporal lobe dose (22.2 Gy versus 24.7 Gy). Compared to IMRT, VMAT techniques required fewer monitor units (755 vs. 478, p ≤ 0.001) but longer optimization times. Treatment delivery times were 6.1 and 10.5 minutes for coplanar and ncIMRT versus 2.9 and 5.0 minutes for coplanar and ncVMAT. In this study, all techniques achieved comparable target coverage. Superior sparing of contralateral optic structures was seen with ncIMRT. The VMAT techniques reduced treatment delivery duration but prolonged plan optimization times, com­pared to IMRT techniques. Technique selection should be individualized, based on patient-specific clinical and dosimetric parameters
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24
    References
    41
    Citations
    NaN
    KQI
    []