New method for time alignment and time calibration of the TOFOR time-of-flight neutron spectrometer at JET.

2021 
The TOFOR time-of-flight (TOF) neutron spectrometer at the Joint European Torus (JET) is composed of 5 start (S1) and 32 stop (S2) scintillation detectors. Recently, the data acquisition system (DAQ) of TOFOR was upgraded to equip each of the 37 detectors with its own waveform digitizer to allow for correlated time and pulse height analysis of the acquired data. Due to varying cable lengths and different pulse processing pathways in the new DAQ system, the 160 (5 · 32) different TOF pairs of start–stop detectors must be time-aligned to enable the proper construction of a summed TOF spectrum. Given the time (energy) resolution required by the entire spectrometer system to measure different plasma neutron emission components, it is of importance to align the detector pairs to each other with sub-nanosecond precision. Previously, the alignment partially depended on using fusion neutron data from Ohmic heating phases of JET experimental pulses. The dependence on fusion neutron data in the time alignment process is, however, unsatisfactory as it involves data one would wish to include in an independent analysis for physics results. In this work, we describe a method of time-aligning the detector pairs by using gamma rays. Given the known geometry and response of TOFOR to gamma rays, the time alignment of the detector pairs is found by examining gamma events interacting in coincidence in both S1–S1 and S1–S2 detector combinations. Furthermore, a technique for separating neutron and gamma events in the different detector sets is presented. Finally, the time-aligned system is used to analyze neutron data from Ohmic phases for different plasma conditions and to estimate the Ohmic fuel ion temperature.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    14
    References
    0
    Citations
    NaN
    KQI
    []