Validation of the Omron HEM-7320-LA, upper arm blood pressure monitor with Intelli Wrap Technology Cuff HEM-FL1 for self-measurement and clinic use according to the European Society of Hypertension International Protocol revision 2010 in the Mexican population

2017 
OBJECTIVE: The aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010. PARTICIPANTS AND METHODS: The evaluation was performed in 39 individuals. The mean age of the participants was 47.9±14 years; systolic BP was 145.2±24.3 mmHg (range: 97-190), diastolic BP was 90.9±12.9 mmHg (range: 68-120), and arm circumference was 30.8±4 cm (range: 25-38.5). RESULTS: The device successfully fulfilled the established criteria of the validation protocol. The device overestimated systolic BP by 0.6±5.7 mmHg and diastolic BP by 2.2±5.1 mmHg. The specially designed cuff HEM-FL1 to cover a broad range of arm circumferences and self-placement fulfilled the requirements of the International Protocol.