Variation of PTV dose distribution on patient size in prostate VMAT and IMRT: a dosimetric evaluation using the PTV dose–volume factor

2014 
Background We propose to use the PTV dose–volume factor (PDVF) to evaluate treatment plans of prostate volumetric modulated arc therapy (VMAT) and intensity modulated radiotherapy (IMRT). Purpose PDVF was used to compare the variation of planning target volume (PTV) coverage between VMAT and IMRT because of weight loss of patient. Materials and methods VMAT and IMRT plans of five patients (prostate volume = 32–86·5 cm 3 ) using the 6 MV photon beams were created with the external contour reduced by depths of 0·5–2 cm to reflect the weight loss. Moreover, integral doses (volume integral of the patient dose) and prostate tumour control probability (TCP) were calculated. Results We found that reduced depth resulted in PDVF decreasing 0·03 ± 4·7 × 10 −4 (VMAT) and 0·04 ± 9·7 × 10 −3 (IMRT) per cm for patients. The decrease of PDVF or degradation of PTV coverage was found more significant in IMRT plans than VMAT with patient size reduction. The integral dose did not change significantly between VMAT and IMRT, while the prostate TCP increased with an increase of reduced depth. Conclusion We concluded that PDVF can be successfully used to evaluate the variation of PTV coverage because of weight loss of patient in prostate VMAT and IMRT. Degradation of PTV coverage in prostate VMAT is less significant than IMRT regarding patient size reduction.