A retrospective dosimetric comparison of TG43 and a commercially available MBDCA for an APBI brachytherapy patient cohort.

2014 
Abstract Purpose To compare dosimetry using a contemporary model based dose calculation algorithm (MBDCA) following TG186 recommendations, and the conventional TG43 method in an 192 Ir high dose rate (HDR) accelerated partial breast irradiation (APBI) patient cohort. Methods Data of 38 APBI patients were studied. Dosimetry for the treatment plans was performed using both the TG43 and TG186 dose calculation methods of the Oncentra Brachy v4.4 treatment planning system (TPS). Analysis included indices of clinical interest for the planning target volume (PTV coverage, dose homogeneity, conformity) as well as dose volume histograms (DVH) for the breast, lung, heart, rib and skin. Significance testing of observed differences between TG43 and TG186 results was carried out and the effect of target location to these differences was studied. Results Statistically significant differences were observed in the values of clinically relevant DVH parameters for the PTV and the organs at risk (OAR), except for the heart. Differences for the PTV are relatively small ( Conclusion While the MBDCA option of the TPS appears to improve dosimetric accuracy, differences from TG43 do not appear to warrant dose prescription changes or treatment protocol amendment..