Dynamic positioning accuracy of a novel multileaf collimator for volumetric modulated arc therapy

2016 
We investigated the dynamic positioning accuracy of Agility (Elekta) for volumetric modulated arc therapy (VMAT). The accuracy of the multileaf collimator (MLC) leaf position during VMAT was evaluated using three different tests: (1) a dynamic multileaf collimator (DMLC) output test with various leaf speeds, and gantry angles; (2) a slit-fence test with and without gantry rotation; and (3) a complicated VMAT plans test with dose distributions compared with measurements using gamma analysis. The DMLC output was within 1.5 % under all test conditions. The agreement between the static and VMAT in the slit-fence test was within 0.5 mm. The pass rate of each complicated VMAT test plan was more than 93.9 % ± 0.36 for gamma analysis. We confirmed the dynamic positioning accuracy of Agility, which during VMAT delivery is within VMAT tolerances. The fastest MLC was found to have the potential to offer clinical advantages, such as high-quality rapid VMAT.