Quantification of respiration-induced esophageal tumor motion using fiducial markers and four-dimensional computed tomography

2016 
Abstract Background and purpose Respiration-induced tumor motion is an important geometrical uncertainty in esophageal cancer radiation therapy. The aim of this study was to quantify this motion using fiducial markers and four-dimensional computed tomography (4DCT). Materials and methods Twenty esophageal cancer patients underwent endoscopy-guided marker implantation in the tumor volume and 4DCT acquisition. The 4DCT data were sorted into 10 breathing phases and the end-of-inhalation phase was selected as reference. We quantified for each visible marker ( n =60) the motion in each phase and derived the peak-to-peak motion magnitude throughout the breathing cycle. The motion was quantified and analyzed for four different regions and in three orthogonal directions. Results The median(interquartile range) of the peak-to-peak magnitudes of the respiration-induced marker motion (left–right/anterior–posterior/cranial–caudal) was 1.5(0.5)/1.6(0.5)/2.9(1.4) mm for the proximal esophagus ( n =6), 1.5(1.4)/1.4(1.3)/3.7(2.6) mm for the middle esophagus ( n =12), 2.6(1.3)/3.3(1.8)/5.4(2.9) mm for the distal esophagus ( n =25), and 3.7(2.1)/5.3(1.8)/8.2(3.1) mm for the proximal stomach ( n =17). Conclusions The variations in the results between the three directions, four regions, and patients suggest the need of individualized region-dependent anisotropic internal margins. Therefore, we recommend using markers with 4DCT to patient-specifically adapt the internal target volume (ITV). Without 4DCT, 3DCTs at the end-of-inhalation and end-of-exhalation phases could be alternatively applied for ITV individualization.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    28
    Citations
    NaN
    KQI
    []