Doses to patients from dental radiology in France.

1989 
In France, a national study was undertaken to estimate both dental radiology practices (equipment and activity) and the associated population collective dose. This study was done in two steps: A nationwide survey was conducted on the practitioner categories involved in dental radiology, and dosimetric measurements were performed on patients and on an anthropomorphic phantom by using conventional dental x-ray machines and pantomographic units. A total of 27.5 x 10(6) films were estimated to have been performed in 1984; 6% of them were pantomographic and 94% were conventional. Most of the organ doses measured for one intra-oral film were lower than 1 mGy (100 mrad); pantomogram dose values were generally higher than intra-oral ones. The collective effective dose equivalent figure was 2,000 person-Sv (2 x 10(5) person rem) leading to a per head dose equivalent of 0.037 mSv (3.7 mrem). The study allowed authors to identify ways to reduce the patient dose in France (e.g., implementing the use of long cone devices and controlling darkroom practices).