High sensitivity cardiac troponin I detection in physiological environment using AlGaN/GaN High Electron Mobility Transistor (HEMT) Biosensors

2018 
Abstract In this study, we report the development of a high sensitivity assay for the detection of cardiac troponin I using electrical double layer gated high field AlGaN/GaN HEMT biosensor. The unique gating mechanism overcomes the drawback of charge screening seen in traditional FET based biosensors, allowing detection of target proteins in physiological solutions without sample processing steps. Troponin I specific antibody and aptamer are used as receptors. The tests carried out using purified protein solution and clinical serum samples depict high sensitivity, specificity and wide dynamic range (0.006–148 ng/mL). No additional wash or sample pre-treatment steps are required, which greatly simplifies the biosensor system. The miniaturized HEMT chip is packaged in a polymer substrate and easily integrated with a portable measurement unit, to carry out quantitative troponin I detection in serum samples with
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    32
    References
    88
    Citations
    NaN
    KQI
    []